ELISA Kit for Angiotensin 1-7 (Ang1-7)

Katalog-Nummer CES085Mi-96

Size : 96T

Marke : Cloud-Clone Corp

Weitere Informationen anfordern

Contact local distributor :


Telefonnummer : +1 850 650 7790



Specificity

This assay has high sensitivity for detection of Angiotensin 1-7 (Ang1-7).
100% cross-reactivity of Angiotensin 1-7 (Ang1-7) was observed among Human, Mouse, Rat.

Recovery

Matrices listed below were spiked with certain level of recombinant Angiotensin 1-7 (Ang1-7) and the recovery rates were calculated by comparing the measured value to the expected amount of Angiotensin 1-7 (Ang1-7) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 84-98 91
EDTA plasma(n=5) 96-103 101
heparin plasma(n=5) 94-105 102

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Angiotensin 1-7 (Ang1-7) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Angiotensin 1-7 (Ang1-7) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Angiotensin 1-7 (Ang1-7) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-104% 84-104% 86-96% 85-95%
EDTA plasma(n=5) 89-103% 99-105% 91-101% 79-101%
heparin plasma(n=5) 78-98% 90-97% 83-94% 84-92%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1 Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
Reagent Diluent 1×300µL Stop Solution 1×6mL
TMB Substrate 1×9mL Instruction manual 1
Wash Buffer (30 × concentrate) 1×20mL

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

Sie könnten auch an folgenden Produkten interessiert sein:



Katalog-Nummer
Beschreibung
Cond.
Preis zzgl. MwSt.
CSB-E14242h-96T
 96Tests