Rat Anti SARS-CoV-2 (COVID) Spike RBD Alpha B.1.1.7 Variant (UK Variant) IgG ELISA Kit

Cat# ER2061

Size : 96T

Brand : FineTest

Request more information

Contact local distributor :


Phone : +1 850 650 7790

This kit is based on indirect ELISA detection method and takes 3h assay time. The microplate provided in this kit has been precoated with antigen. HRP-antibody is used as detection antibody. Add standard and properly diluted sample into relevant well respectively. After incubation, wash unbound components. Add HRP-detection antibody, then it binds with Anti SARS-CoV-2 UK Variant IgG bound to precoated antibody. Wash unbound components and add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Calculate the concentration of Anti SARS-CoV-2 UK Variant IgG in the sample by plotting standard curve. The concentration of the target substance is proportional to the OD450 value.

Catalogue No.: 
ER2061

Alias: 
SARS-CoV-2 UK Variant IgG ELISA Kit

Species: 
Rat

Range: 
1.56-100ng/ml

Sensitivity: 
0.938ng/ml

  • SPECIFICATIONS
  • FIGURES
  • CONDITIONS
Product Name
Rat Anti SARS-CoV-2 (COVID) Spike RBD Alpha B.1.1.7 Variant(UK Variant) IgG ELISA Kit

Alias
SARS-CoV-2 UK Variant IgG ELISA Kit

Catalogue No.
ER2061

Size
48T/96T

Species
Rat

Sample Type
Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples

Detection Method
Indirect ELISA Sandwich

Detection Wavelength
OD450

Reaction Duration
3 hours

Range
1.56-100ng/ml

Sensitivity
0.938ng/ml

Storage
2-8°C(Sealed), Don't cryopreserve.

Specificity
Specifically binds with Anti SARS-CoV-2 UK Variant IgG , no obvious cross reaction with other analogues.

ELISA Kit Components
Kit Components Item Size(48T) Size(96T) Storage Condition for Opened Kit
E001 ELISA Microplate(Dismountable) 8×6 8×12 Put the rest strips into a sealed foil bag with the desiccant. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E002 Lyophilized Standard 1vial 2vial Put the rest standards into a desiccant bag. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E003 HRP Antibody(Concentrated, 100X) 60ul 120ul 2-8°C (Avoid Direct Light)
E024 TMB Substrate 5ml 10ml
E039 Sample Dilution Buffer 10ml 20ml 2-8°C
E040 Antibody Dilution Buffer 5ml 10ml
E026 Stop Solution 5ml 10ml
E038 Wash Buffer(Concentrated, 25X) 15ml 30ml
E006 Plate Sealer 3 pieces 5 pieces  
E007 Product Description 1 copy 1 copy

Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water

Assay Procedure Summary
Elisa实验原理图
  • Step 1: Wash the plate twice before adding the standard and sample.
  • Step 2: Add 100ul standard or sample into each well, seal the plate and statically incubate for 90 minutes at 37°C.
  • Washing: Wash the plate three times and immerse for 1min each time.
  • Step 3: Add 100ul HRP-antibody working solution, seal the plate and statically incubate for 30 minutes at 37°C.
  • Washing: Wash the plate five times and immerse for 1min each time.
  • Step 4: Add 90ul TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
  • Step 5: Add 50ul stop solution. Read at 450nm immediately and calculate.

Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(ng/ml) OD-1 OD-2 Average
0 0.1 0.102 0.101
1.562 0.166 0.17 0.168
3.125 0.296 0.304 0.3
6.25 0.45 0.464 0.457
12.5 0.715 0.735 0.725
25 0.956 0.984 0.97
50 1.732 1.782 1.757
100 2.306 2.372 2.339
ER2061 Standard Curve Image

Recovery

Add a certain amount of Anti SARS-CoV-2 UK Variant IgG into the sample. Calculate the recovery by comparing the measured value with the expected amount of Anti SARS-CoV-2 UK Variant IgG in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 86-104 91
EDTA plasma(n=10) 87-99 94
Heparin plasma(n=10) 95-101 97

Linearity

Dilute the sample with a certain amount of Anti SARS-CoV-2 UK Variant IgG at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 86-105% 88-101% 92-100%
EDTA plasma(n=10) 86-104% 93-100% 88-99%
Heparin plasma(n=10) 93-102% 87-99% 88-99%

Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (ng/ml) 2.94 12.13 47.6 3.14 11.84 48.71
Standard deviation 0.15 0.71 2.46 0.15 0.67 2.34
CV(%) 5.27 5.84 5.17 4.68 5.62 4.8

Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months
Average(%) 80 95-100