Human COVID-19 Spike protein (S2) IgA ELISA Kit

Cat# EH5017

Size : 96T

Marca : FineTest

Richiedere ulteriori informazioni

Contatta il distributore locale :


Telefono : +1 850 650 7790

This kit is based on indirect ELISA detection method and takes 4h assay time. The microplate provided in this kit has been precoated with antigen. Biotinylated antibody is used as detection antibody. Add standard and properly diluted sample into relevant well respectively. After incubation, wash unbound components. Add biotinylated detection antibody. Then, it binds with COVID-19 Spike protein (S2) IgA bound to precoated antigen. Wash unbound components and add HRP-Streptavidin Conjugate (SABC). Wash unbound components again and add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Calculate the concentration of COVID-19 Spike protein (S2) IgA in the sample by plotting standard curve. The concentration of the target substance is proportional to the OD450 value.

Catalogue No.: 
EH5017

Alias: 
COVID-19 Spike protein (S2) IgA ELISA Kit

Species: 
Human

Range: 
0.391-25ng/ml

Sensitivity: 
0.234ng/ml

  • SPECIFICATIONS
  • FIGURES
  • CONDITIONS
Product Name
Human COVID-19 Spike protein (S2) IgA ELISA Kit

Alias
COVID-19 Spike protein (S2) IgA ELISA Kit

Catalogue No.
EH5017

Size
48T/96T

Species
Human

Sample Type
Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples

Detection Method
Indirect ELISA

Detection Wavelength
OD450

Reaction Duration
2 hours

Range
0.391-25ng/ml

Sensitivity
0.234ng/ml

Storage
2-8°C(Sealed), Don't cryopreserve.

Specificity
Specifically binds with COVID-19 Spike protein (S2) IgA , no obvious cross reaction with other analogues.

Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water

Assay Procedure Summary
Elisa实验原理图

Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(ng/ml) OD-1 OD-2 Average
0 0.086 0.091 0.088
0.391 0.172 0.181 0.176
0.781 0.241 0.253 0.246
1.562 0.374 0.393 0.382
3.125 0.653 0.686 0.666
6.25 1.046 1.099 1.067
12.5 1.602 1.684 1.635
25 2.15 2.26 2.194
EH5017 Standard Curve Image

Recovery

Add a certain amount of COVID-19 Spike protein (S2) IgA into the sample. Calculate the recovery by comparing the measured value with the expected amount of COVID-19 Spike protein (S2) IgA in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 85-101 95
EDTA plasma(n=10) 86-100 91
Heparin plasma(n=10) 86-104 95

Linearity

Dilute the sample with a certain amount of COVID-19 Spike protein (S2) IgA at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 87-101% 83-101% 86-92%
EDTA plasma(n=10) 85-104% 85-96% 80-86%
Heparin plasma(n=10) 94-104% 82-96% 82-92%

Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (ng/ml) 0.84 3.02 12.55 0.82 3.24 13.38
Standard deviation 0.04 0.15 0.4 0.04 0.17 0.69
CV(%) 5.21 5.08 3.22 4.69 5.12 5.12

Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months
Average(%) 80 95-100